Didymium Glass Filter

Description Didymium glass filter, solid filter, certified
Use Assessment of the photometric accuracy in the UV region (270 nm to 340 nm)
Material Glass BG 36 of SCHOTT Glaswerke

666-F7 Didymium glass filter

E 015 Kalibrierstandards

The F7 marked didymium glass filter is actually a standard to check for wavelength accuracy in the UV region (see page 5), but it can also be used for checking the photometric accuracy.

Checking the photometric accuracy in the UV range is possible at the following wavelengths: 270 nm, 280 nm, 300 nm, 320 nm and 340 nm. As a standard, the thickness of the filter is manufactured in a way that it will have an absorbance of approx. 0.5 A at a wavelength of 340 nm. Absorbance values will increase with decreasing wavelengths up to ca. 1.5 A at 270 nm.

E_Didymiumglas-Filter Typical spectrum of a Didymium glass filter between 270 nm and 370 nm
Catalogue-
No.
Description Material Thick-
ness mm
Nominal Absorbance Absorbance
value
at nm
666-F7 Didymium Glass Filter,
Certified
BG 36 ca. 1,2 depending on
wavelength
270, 280, 300, 320, 340